Wall Mount UV-Fluorescence Oil in Water
Monitor FPM 605
The product
brochure will be available for download soon!
Technical Data
- Method of operation: UV fluorescence detector
- Dimensions (W x D x H): 430 mm x 480 mm x 270 mm
- Weight approx.: 18 kg (40 lb.)
- Detection range: 1ppm - 1000 ppm (detection limits depend
on target analyte and matrix)
- Plumbing requirements: Feed in: approx. 200 ml/min, 2"
unrestricted drain
- Response time: < 12 seconds, depending on actual flow
through sample cell
- Operating temperature: 5 - 45°C, for extreme temperature
conditions contact manufacturer
- Ambient temperature: 5-45°C (41-110°F)
- Sample temperature: Max 60°C
- Calibration: Multi point (min. 2 point)
- Zero drift: <2.0% FSD / 24h
- Span drift: <2.0% FSD / 24h
- Linearity: 1,5%
- Measuring ranges (ppm): 0-100 and 0-1.000, alt. 0-200 and
0-2000ppm
- Analog outputs: 0-10VDC and 4-20mA; RS 232 data out optional
- Display : 3 1/2-digit, red LED, 6-digit engineering display
optional
- Sample flow: approx. 250 cc/min through flow cell
- Zero and span adjust: Manual on front panel
- Routine maintenance: Lamp once every 20,000 hours, routine
cleaning of optics, flow cell and sample containing plumbing
- Air consumption for air curtain: approx. 1liter per minute
(built it air supply optional)
- Temperature control: Optional for cold weather conditions,
digital PID-type controller
- Power requirements: 115VAC/60Hz, 150 W, or 230VAC/50Hz
Available Options
- ENGA 65 6-digit engineering units display, 0-10.000 ppm with
RS 232 data output, no range change necessary over the span of
three 10VDC analog decades (between 0 to 1000ppm) Range changing
is only required when an analog strip chart recorder is used
- AMU 65 Automatic range change
- CIM 65 RS 232 data output, not necessary with ENGA 65
- EAP 65 Air purge kit to meet class I Division II (explosion
proofing)
- DBP 65 Water overflow sampling and de-bubbeling system. Accurate,
bubble free sample flow regulation due to height adjusted mounting
of the system besides the right side of the monitor next to sample
inlet
- PAS 65 External hydrocarbon free air supply with pump and
air purifier for air curtain
- RCA 65 0-20mA analog output instead of 4-20mA
- RCC 65 Remote control range change
- RCI0 65 0-20 mA analog output, galvanically isolated
- RCI4 65 4-20 mA analog output, galvanically isolated